Western Blotting Workflow
Step-by-Step Visualized Ultra-Fast Western Blot
Experimental Procedure
Objective:
- To experience the Western Blot experimental procedure and understand the value of step-by-step visualization in protein immunoblotting technology.
- To familiarize with the functions of the imaging analyzer in result recording, image optimization, band quantification, and quantitative normalization.
Experimental Samples:
-
- Marker Standard 1: Precision Plus Protein™ All Blue Standards --- Tube labeled "B"
- Marker Standard 2: Precision Plus Protein™ Unstained Standards --- Tube labeled "U"
- Hela cell lysate: Original --- Tube labeled "F1"
- Hela cell lysate: 1:2 dilution --- Tube labeled "F2"
- Hela cell lysate: 1:4 dilution --- Tube labeled "F3"
- Hela cell lysate: 1:8 dilution --- Tube labeled "F4"
- Hela cell lysate: 1:8 dilution --- Tube labeled "F5"
-
Preparation of test samples: Starting from the provided F1 (already mixed with Loading Buffer), perform serial dilutions as shown in the table below:
Tube Label Content F2 15uL Loading Buffer + 15uL F1 F3 15uL Loading Buffer + 15uL F2 F4 15uL Loading Buffer + 15uL F3 F5 15uL Loading Buffer + 15uL F4 (Note: UTF is untreated fish meat extract.)
- Place F1, F2, F3, F4, and F5 in a water bath for high-temperature denaturation at 95°C for 5 minutes. Afterward, remove the samples and arrange them in the loading order.
Gel Electrophoresis Steps
Electrophoresis - Precast Gel Group
- Take out a precast gel (4-20% Stain-Free Mini-PROTEAN® TGX™ 10-well gel), remove the sealing tape at the bottom, assemble it into the Mini Tetra electrophoresis module, and then remove the comb.
- If running only one gel, use a buffer dam or a discarded TGX gel with sealing tape to seal the opposite side of the module.
- Fill the inner chamber of the assembled electrophoresis module with electrophoresis buffer (1x TGS Buffer) and check for leaks.
- Remove the comb, gently rinse the sample wells with a plastic pipette (or 1 mL tip) to remove bubbles and straighten the wells. Load the samples and standards in the order shown below:

Well Volume Sample Name 1 10 ul All Blue Standards ----B 2 10 ul Unstained Standards ---U 3 10 ul Hela-1 (F1) 4 10 ul Hela -2 (F2) 5 10 ul Hela -3 (F3) 6 10 ul Hela -4 (F4) 7 10 ul Hela -5 (F5) 8 10 ul Hela -5 (F5) 9 10 ul Unstained Standards ---U 10 10 ul All Blue Standards ----B - Place the loaded electrophoresis module into the buffer tank and fill the tank with the same buffer to the appropriate level.
- Set the electrophoresis conditions to 250-300 V constant voltage for 15-20 minutes. If possible, place the electrophoresis tank on a stirrer with a stir bar for temperature equilibration.
- After electrophoresis, remove the gel from the cassette, place it in water, and cut a corner for marking. Use the W5000 imaging system to capture the blotting results.
- Prepare for the transfer step.
Protein Blotting Experimental Steps --- Transfer Section
Using Transfer Instrument for Rapid Transfer:
Mini Transfer Packs Midi Transfer Packs
The anode stack is placed on the right end (Bottom+). Use the right finger hole for removal.


- Place the membrane and lower consumables (anode) on the transfer slot.
- Use a specialized roller to remove bubbles.
- For 1 small or medium gel, place it in the center of the transfer slot.
- For 2 small gels, place them side by side without overlapping.
- Place the electrophoresed gel on top of the membrane.
- No gel equilibration is required before transfer.
- Use a specialized roller to remove bubbles.
- Place the upper consumables (cathode) on top of the gel.
- Use a specialized roller to remove bubbles.
- Close the transfer lid and lock it by rotating. Insert the correctly assembled transfer cassette into the Trans-Blot Turbo system to begin the experiment.
- After powering on, select transfer instrument from the home screen's lower right corner to access preset programs. Choose 2 MINI OR 1 MIDI GEL, which sets the transfer conditions to [Mixed MW, 25V, 2.5A, 7 minutes]. Press "RUN" to start.
W5000 Multi-Functional Imaging System Demo Standard Operating Procedure
--- Immunodetection Section
- After the transfer process, remove the PVDF membrane and check the transfer efficiency by verifying if the pre-stained protein standards have transferred onto the membrane.
- Cut a corner of the membrane for marking.
- Place the transferred PVDF membrane (gel-contact side up) into blocking buffer and incubate on a shaker for 1 hour.
- Replace with 50 mL of primary antibody blocking solution and incubate for 1 hour at room temperature.
- Rinse with wash buffer 3 times, 5 minutes each, on a shaker.
- Replace with 50 mL of secondary antibody blocking solution + Strep-Tactin HRP and incubate on a shaker for 15 minutes.
- Repeat the rinsing steps 3 times, 5 minutes each, on a shaker.
- Perform chemiluminescent detection using freshly prepared Clarity ECL substrate solution.
Reaction mixture =0.5 mL A (white bottle) + 0.5 mL B (black bottle).
- Mix well and apply evenly to the membrane surface (use a pipette tip for even spreading). Allow the reaction to proceed for 1--3 minutes, then remove excess liquid without letting the membrane dry.
- Capture the blotting results using the W5000 imaging system: Use the Chemi application in the Analysis Software with multiple cumulative exposure times (10 seconds, 4 images).
- Save the results file and analyze the data using the Analysis Software.