model:FT-700/FT--700plus/FT--800/FT--5000
application:It can quickly and accurately detect nucleic acid, protein and cell solutions, and is equipped with a colorimetric dish mode to detect the concentration of bacteria and other culture media. It can be used in clinical disease diagnosis, blood transfusion safety, forensic identification, environmental microbial detection, food safety detection, molecular biology research and other fields.
|
Product model
|
FT--700
|
FT--700plus
|
FT--800
|
FT--5000
|
|
Illuminant
|
LED
|
xenon flashtube
|
|
Detection technique
|
The "3" optical path detection mode is adopted
|
The "4" optical path detection mode is adopted
|
|
Sample concentration
|
The sample does not need to be diluted, and can measure more than 150 times of the conventional UV-visible photometer;
|
|
Fluorescence function
|
-
|
The pg concentration of dsDNA can be detected with fluorescence quantitative reagent
|
-
|
|
Built-in printer
|
Print reports directly from the built-in printer
|
|
Operating system
|
Android operating system, 7 inch capacitive touch screen, no computer online, single can be detected.
|
|
Detection system
|
OD600 Optical path detection system
|
|
Data processing
|
Image JPG and table Excel storage format
|
|
Optical path
|
0.05mm(High concentration measurement);0.2mm、1.0mm(General concentration measurement)
|
0.02mm、0.05mm(High concentration measurement);0.2mm、1.0mm(General concentration measurement
)
|
|
Sample volume requirement
|
0.5-2μl
|
|
Base measurement volume
|
0.5~2ul
|
|
Wavelength range
|
260nm、280nm
|
190-850nm
|
|
Wavelength accuracy
|
±1nm
|
|
Wavelength reproducibility
|
< ±0.2nm
|
|
OD600 absorbance range
|
0~6.000Abs
|
|
OD600 absorbance stability
|
[0,3)≤0.5%;[3,4)≤2%;
|
|
OD600 absorbance accuracy
|
[0,2)≤0.005A;[2,3)≤1%;[3,4)≤2%;
|
|
OD600 absorbance repeatability
|
[0,3)≤0.5%;[3,4)≤2%;
|
|
Detector
|
Ultraviolet silicon photocell
|
HAMAMATSU Uv enhanced type CMOS Linear array sensor;
|
|
Nucleic acid range
|
0~15000 ng/µL dsDNA
|
0~27500 ng/µL dsDNA /0.06~820 mg/mL BSA
|
|
Fluorescence sensitivity
|
-
|
DsDNA:0.5pg/ul
|
DsDNA:0.5pg/ul
|
-
|
|
Fluorescence linearity
|
-
|
≤1.5%
|
≤1.5%
|
-
|
|
Light absorption range
|
0.04-300A(10mm)
|
0.02-300A(10mm)
|
|
Absorbance accuracy
|
0.003Abs(0.2mm optical path)
|
0.002Abs(0.2mm optical path)
|
|
Absorbance trueness
|
±1%(7.332Abs at 260nm)
|
|
spectral resolution
|
-
|
-
|
<1nm
|
|
Photometric accuracy
|
±0.005A or ±1%
|
|
Sample wavelength scanning
|
-
|
-
|
190~850nm
|
|
Noise level
|
0.002Arms(0A,260nm) ~ 0.005A(0A,260nm)
|
|
Sample base material
|
Quartz fiber and stainless steel
|
|
Detection time
|
<6s
|
<5s
|
|
Standby power consumption
|
≤5W
|
|
Data output mode
|
USB
|
|
Illuminant
|
LED
|
|
Power adapter
|
24V DC
|
|
Power dissipation
|
≤ 25W
|
|
Overall dimensions W*D*H
|
200×260×165
|
200×260×165
|
200×260×165
|
200×260×165
|
|
weight
|
4.5Kg
|
4.5Kg
|
5Kg
|